GM700S

GM700S

Measuring blood sugar is now simple with GM700S new before-and-after meal markers on top of its compact and easy-to-use design.
Wide HCT range

Measures glucose value of extreme hematocrit range, like neonate, elder, anemia patient.

Meal markers

Before-and-after meal markers allow you to understand how food affects your blood glucose.

Alternative blood sampling site

Capillary, venous, arterial and neonatal whole blood are applicable.

Auto coding

Eliminates the needs for manual coding and delivers better accuracy than no coding.

Proprietary Rightest™ strip

With this Validus Technology™, Rightest™ strips always deliver high accuracy and precision in readings.

Safe & Hygienic

The innovative side-insert test strip allows users to avoid contact with the blood sample

How it Works

1. Insert a GS700 test strip in the socket of the meter. The calibration will be done automatically.

2. Place your blood sample against the edge of sample entry of the strip.

3. After 5 seconds, the test result will appear.

4. Rotate and pull up the test strip from the meter.

Play Video

Specification

Test stripGS700
CodingAuto coding calibration
Enzyme / Measurement technologyGDH / Electrochemical sensor
Blood sample sourceCapillary, venous, arterial and neonatal whole blood
Minimum sample volume0.75 microliters  (0.00075c.c.)
HCT range

Glucose concentrations ≦ 200mg/dL, HCT range 20 – 70%

Glucose concentrations > 200mg/dL, HCT range 20 – 60%

Measuring range10 – 600mg/dL (0.6 – 33.3 mmol/L)
Meter dimensions (L x W x H)82.0 x 45.0 x 15.5 mm (3.22 x 1.77 x 0.61 inch)
Display area (L x W)34.0 x 27.5 mm (1.09 x 1.34 inch)
Weight (with battery)43.0 ± 5 g
Power supply1 CR2032 battery
Memory capacity500 blood glucose test results with date and time
Operating temperature6 – 44℃ (43 – 111°F)
Operating relative humidity10 – 90%
Meter storage conditions-10 – 60°C (14 – 140°F)
Test strip storage conditions4 – 30°C (39 – 86°F), < 90% relative humidity
Accuracy

95% of the measured glucose values shall fall within either

Glucose concentrations < 5.5mmol/L (100mg/dL), difference within ± 0.83mmol/L (15mg/dL)

Glucose concentrations ≥ 5.5mmol/L (100mg/dL), bias within ± 15%

Precision

Glucose concentrations < 5.5mmol/L (100mg/dL), SD < 0.28mmol/L (5mg/dL)

Glucose concentrations ≥ 5.5mmol/L (100mg/dL), CV < 5%

Fully comply with ISO 15197:2013 (For other received credentials, please contact us.)