Accurate measurement in 5 seconds.
![](https://bionime.com.my/wp-content/uploads/2020/04/website_icon-04.png)
Only 0.75 microliters (0.00075c.c.) of blood sample required.
![](https://bionime.com.my/wp-content/uploads/2020/04/website_icon-05.png)
Eliminates the needs for manual coding and delivers better accuracy than no coding.
![](https://bionime.com.my/wp-content/uploads/2020/04/website_icon-09.png)
Easy to view in a dark environment.
![](https://bionime.com.my/wp-content/uploads/2020/04/website_icon-06.png)
Large display and text is easy to read.
![](https://bionime.com.my/wp-content/uploads/2020/04/website_icon-08.png)
Data transmission via PC linkage makes blood glucose control easier.
How it Works
![](https://bionime.com.my/wp-content/uploads/2020/04/01_GM550-Step.jpg)
1. Insert a GS550 test strip in the socket of the meter. The calibration will be done automatically.
![](https://bionime.com.my/wp-content/uploads/2020/04/02_GM550-Step.jpg)
2. Place your blood sample against the edge of sample entry of the strip.
![](https://bionime.com.my/wp-content/uploads/2020/04/03_GM550-Step.jpg)
3. After 5 seconds, the test result will appear.
![](https://bionime.com.my/wp-content/uploads/2020/04/04_GM550-Step_en.jpg)
4. Rotate and pull up the test strip from the meter.
![](https://bionime.com.my/wp-content/uploads/2020/04/video_photo_1.jpg)
Specification
Test strip | GS550 |
Coding | Auto coding calibration |
Enzyme / Measurement technology | GOD / Electrochemical sensor |
Blood sample source | Capillary whole blood |
Minimum sample volume | 0.75 microliters (0.00075c.c.) |
HCT range | 30 – 57% |
Measuring range | 10 – 600mg/dL (0.6 – 33.3 mmol/L) |
Meter dimensions (L x W x H) | 90.6 x 46.0 x 16.5 mm (3.56 x 1.81 x 0.64 inch) |
Display area (L x W) | 46.6 x 33 mm (1.83 x 1.32 inch) |
Weight (with battery) | 53.0 ± 5 g |
Power supply | 2 CR2032 batteries |
Memory capacity | 500 blood glucose test results with date and time |
Operating temperature | 10 – 40°C (50 – 104°F) |
Operating relative humidity | 10 – 90% |
Meter storage conditions | -10 – 60°C (14 – 140°F) |
Test strip storage conditions | 4 – 30°C (39 – 86°F), < 90% relative humidity |
Accuracy | 95% of the measured glucose values shall fall within either Glucose concentrations < 5.5mmol/L (100mg/dL), difference within ± 0.83mmol/L (15mg/dL) Glucose concentrations ≥ 5.5mmol/L (100mg/dL), bias within ± 15% |
Precision | Glucose concentrations < 5.5mmol/L (100mg/dL), SD < 0.28mmol/L (5mg/dL) Glucose concentrations ≥ 5.5mmol/L (100mg/dL), CV < 5% |