Accurate measurement in 5 seconds.
Only 0.75 microliters (0.00075c.c.) of blood sample required.
Eliminates the needs for manual coding and delivers better accuracy than no coding.
Easy to view in a dark environment.
Large display and text is easy to read.
Data transmission via PC linkage makes blood glucose control easier.
How it Works
1. Insert a GS550 test strip in the socket of the meter. The calibration will be done automatically.
2. Place your blood sample against the edge of sample entry of the strip.
3. After 5 seconds, the test result will appear.
4. Rotate and pull up the test strip from the meter.
Specification
Test strip | GS550 |
Coding | Auto coding calibration |
Enzyme / Measurement technology | GOD / Electrochemical sensor |
Blood sample source | Capillary whole blood |
Minimum sample volume | 0.75 microliters (0.00075c.c.) |
HCT range | 30 – 57% |
Measuring range | 10 – 600mg/dL (0.6 – 33.3 mmol/L) |
Meter dimensions (L x W x H) | 90.6 x 46.0 x 16.5 mm (3.56 x 1.81 x 0.64 inch) |
Display area (L x W) | 46.6 x 33 mm (1.83 x 1.32 inch) |
Weight (with battery) | 53.0 ± 5 g |
Power supply | 2 CR2032 batteries |
Memory capacity | 500 blood glucose test results with date and time |
Operating temperature | 10 – 40°C (50 – 104°F) |
Operating relative humidity | 10 – 90% |
Meter storage conditions | -10 – 60°C (14 – 140°F) |
Test strip storage conditions | 4 – 30°C (39 – 86°F), < 90% relative humidity |
Accuracy | 95% of the measured glucose values shall fall within either Glucose concentrations < 5.5mmol/L (100mg/dL), difference within ± 0.83mmol/L (15mg/dL) Glucose concentrations ≥ 5.5mmol/L (100mg/dL), bias within ± 15% |
Precision | Glucose concentrations < 5.5mmol/L (100mg/dL), SD < 0.28mmol/L (5mg/dL) Glucose concentrations ≥ 5.5mmol/L (100mg/dL), CV < 5% |