GE Max Series

GE Max Series

With intuitive design and Bluetooth transmission, your blood glucose measurement can be as easy as snapping fingers!
Bluetooth transmission

Transfers test results to Rightest CARE App wirelessly.

Wide HCT range

Measures glucose value of extreme hematocrit range, like neonate, elder, anemia patient.

Alternative blood sampling site

Capillary, venous, arterial and neonatal whole blood are applicable.

Auto coding

Eliminates the needs for manual coding and delivers better accuracy than no coding.

Status marker

Intuitive markers for time status, exercise and ‘before’ or ‘after’ a meal.

Ketone alert

Easy ketone test reminder setting.

Marker

Wake up
Exercise
Bedtime
Midnight
Before Meal
After Meal

Glucose Variability

Hyper
Normal
Hypo

Say Goodbye to Logbook

Manage blood glucose and maintain health easily

How it Works

  1. Insert a Max test strip in the socket of the meter. The calibration will be done automatically.

2. Place your blood sample against the edge of sample entry of the strip.

3. After 5 seconds, the test result appears.

Rotate and pull up the test strip from the meter.

Play Video

Specification

Test stripMax Strip
CodingCoding is not required
Enzyme / Measurement technologyGDH / Electrochemical sensor
Blood sample sourceCapillary, venous, arterial and neonatal whole blood
Minimum sample volume0.75 microliters  (0.00075c.c.)
HCT range10 – 70%
Measuring range0.6 – 33.3mmol/L (10 – 600mg/dL)
Meter dimensions (L x W x H)82.0 x 50.0 x 15.5 mm (3.23 x 1.97 x 0.61 inch)
Display area (L x W)40.7 x 40.2 mm (1.60 x 1.58 inch)
Weight (with battery)59.0 ± 5 g
Power supply1 CR2032 battery (Max)    |   2 CR2032 battery (Max Plus )
Memory capacity500 blood glucose test results with date and time
Operating temperature6 – 44°C (43 – 111°F)
Operating relative humidity10 – 90%
Meter storage conditions-10 – 60°C (14 – 140°F)
Test strip storage conditions4 – 30°C (39 – 86°F), < 90% relative humidity
Bluetooth version (For Max Plus)4.0
Accuracy

95% of the measured glucose values shall fall within either

Glucose concentrations < 5.5mmol/L (100mg/dL), difference within ± 0.83mmol/L (15mg/dL)

Glucose concentrations ≥ 5.5mmol/L (100mg/dL), bias within ± 15%

Precision

Glucose concentrations < 5.5mmol/L (100mg/dL), SD < 0.28mmol/L (5mg/dL)

Glucose concentrations ≥ 5.5mmol/L (100mg/dL), CV < 5%

Fully comply with ISO 15197:2013 (For other received credentials, please contact us.)