GM700

GM700

Professional Medical Specifications with Optimized Functions

The new glucose dehydrogenase reagent makes your medical needs possible in a more comprehensive way—enabling you to detect a variety of blood samples.
Wide HCT range

Measures glucose value of extreme hematocrit range, like neonate, elder, anemia patient.

Replaceable Strip Holder

Eliminates blood samples contamination through convenient strip holder exchange.

Insufficient blood sample warning

Insufficient blood sample will result in an error message 'Er4'.

Alternative blood sampling site

Capillary, venous, arterial and neonatal whole blood are applicable.

Auto coding

Eliminates the needs for manual coding and delivers better accuracy than no coding.

Easy-to-read display

Large display and text.

How it Works

1. Insert a GS700 test strip in the socket of the meter. The calibration will be done automatically.

2. Place your blood sample against the edge of sample entry of the strip.

3. After 5 seconds, the test result will appea

4. Rotate and pull up the test strip from the meter.

Play Video

Specification

Test stripGS700
CodingAuto coding calibration
Enzyme / Measurement technologyGDH / Electrochemical sensor
Blood sample sourceCapillary, venous, arterial and neonatal whole blood
Minimum sample volume0.75 microliters  (0.00075c.c.)
HCT range

Glucose concentrations ≦ 200mg/dL, HCT range 20 – 70%

Glucose concentrations > 200mg/dL, HCT range 20 – 60%

Measuring range10 – 600mg/dL (0.6 – 33.3 mmol/L)
Meter dimensions (L x W x H)99.0 x 46.0 x 17.5 mm (3.89 x 1.81 x 0.68 inch)
Display area (L x W)52.6 x 32.0 mm (2.07 x 1.25 inch)
Weight (with battery)57.0 ± 5 g
Power supply1 CR2032 battery
Memory capacity1,000 blood glucose test results with date and time
Operating temperature6 – 44℃ (43 – 111°F)
Operating relative humidity10 – 90%
Meter storage conditions-10 – 60°C (14 – 140°F)
Test strip storage conditions4 – 30°C (39 – 86°F), < 90% relative humidity
Accuracy

95% of the measured glucose values shall fall within either

Glucose concentrations < 5.5mmol/L (100mg/dL), difference within ± 0.83mmol/L (15mg/dL)

Glucose concentrations ≥ 5.5mmol/L (100mg/dL), bias within ± 15%

Precision

Glucose concentrations < 5.5mmol/L (100mg/dL), SD < 0.28mmol/L (5mg/dL)

Glucose concentrations ≥ 5.5mmol/L (100mg/dL), CV < 5%

Fully comply with ISO 15197:2013 (For other received credentials, please contact us.)